Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
-
View Affiliations Hide AffiliationsKaisu Rantakokko-Jalavakaisu.rantakokko-jalava tyks.fi
-
View Citation Hide Citation
Citation style for this article: . Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019. Euro Surveill. 2019;24(22):pii=1900298. https://doi.org/10.2807/1560-7917.ES.2019.24.22.1900298 Received: 14 May 2019; Accepted: 30 May 2019
- Previous Article
- Table of Contents
- Next Article
Abstract
Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.